Automatic Mode Switch (AMS) Causes Less Synchronization
Author(s):
Abstract:
Introduction
Cardiac resynchronization devices are part of modern heart failure management. After implantation, we analyze and program devices in an attempt to ensure their success. Biventricular pacing should be 98% or more for the lowest mortality and best symptom improvement.Case Presentation
In this case series, we present a combination of far field sensing and automatic mode switching (AMS) in six patients. It is found that this combination causes ventricular sensing (VS) episodes with wide QRS and no synchronization. We turn off the AMS and alleviate the problem.Conclusions
Switching AMS off may increase biventricular pacing in some patients.Language:
English
Published: