Abstract:
Introduction
To evaluate the accuracy of transcutaneous bilirubin measurement in a large population of newborn infants, before and during the phototherapy.
PATIENTS AND
Methods
A single Bilicheck instrument was used for transcutaneous measurements. A photo-opaque patch was positioned over the measurement site prior to starting phototherapy. Transcutaneous bilirubinometry was conducted on an unpatched area of the forehead skin and on the nearby site covered by the photo-opaque patch. Readings were obtained from patched and unpatched areas and simultaneous total serum bilirubin concentrations were compared.
Results
We studied 134 term and 36 preterm newborns. Pre-phototherapy measurements showed a strong correlation (r: 0.929, P
Conclusion
BiliCheck can be safely used for the evaluation of bilirubin levels in preterm and term newborn infants under phototherapy. BiliCheck is slightly less reliable among preterm newborns.
Language:
English
Published:
Archives of Iranian Medicine, Volume:19 Issue: 5, May 2016
Page:
323
magiran.com/p1543329  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!