Impacts of Increasing the Number of IMRT Beams on Heart’s Dose Distribution in Left Breast Irradiation: Dosimetric Study

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Introduction
The main objective of this study was to assess the impacts of an increasing the number of IMRT beams on cardiac dose distribution in left-sided breast irradiation so that we can reduce the heart’s mean dose  up to clinically acceptable level.
Material and Methods
For this study 107 female patients, diagnosed with left-sided breast cancer were selected retrospectively. In 107 patients, there were 52 patients of chest wall irradiation including supra-clavicular fossa, while 22 patients were of breast-conserving surgery excluding supra-clavicular fossa and internal mammary lymph nodes, and 33 patients were of chest wall irradiation including internal mammary lymph nodes and supra-clavicular fossa. Exclusion criteria were previous history of left-sided breast radiation therapy, uncommon fractionated dose delivered in past, and indication of palliative radiation therapy. Intensity modulated radiotherapy plans were generated using 7, 9 and11 beams for each patient and the prescribed dose was 40.05 Gy in 15 fractions (2.67 Gy /fraction) for the targets.
Results
Heart: V5Gy(cc): This was a low-dose volume of our study in which the 11-bIMRT technique yielded better result as compared to 9- and 7-bIMRT. Maximum and minimum values of V5 were found 539.60cc in 9-bIMRT and 141.32cc in 11-bIMRT techniques respectively.V25Gy(cc): The maximum value of V25Gy was found 41.73cc in 7-bIMRT technique, while the lowest value was 0.29cc in 11-bIMRT. The IMRT technique with 11 beams showed comparatively better result on this parameter as well as 3-5cc volume of V25Gy was spared. Mean dose (Gy): Maximum value of mean dose was found 8.51Gy in 7-bIMRT while it was 6.53Gy in 11-bIMRT technique.   
Conclusion
The study indicates that increasing the number of IMRT beams reduces heart’s high-dose volume and improves the quality of treatment plans. It is judicious to use 11-bIMRT technique in left-sided breast irradiation as it produces clinically acceptable mean heart dose.
Language:
English
Published:
Iranian Journal of Medical Physics, Volume:19 Issue: 3, May-Jun 2022
Pages:
145 to 153
magiran.com/p2445666  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!