An Investigation of the Dose Distribution from LDR Ir-192 Wires in the Triangular Implants of the Paris System using Polymer Gel Dosimetry

Message:
Abstract:
Introduction
Polymer gels are modern dosimeters providing three dimensional dose distributions. These dosimeters can be used in brachytherapy in which the tumor dimension is relatively small and the dose gradient is high. In this study, the ability of the MAGICA polymer gel was investigated for assessing the absolute dose values as well as the dose distribution of low dose rate (LDR) Ir-192 wires in interstitial brachytherapy based in triangular implants of the Paris system.
Material And Methods
A suitable phantom was made from Perspex. Glass tubes were used as the external tubes for holding the Ir-192 wires in the phantom. The MAGICA polymer gel was made and placed in the phantom. The phantom and the calibration tubes were irradiated using LDR Ir-192 wires and a Co-60 teletherapy unit respectively. They were subsequently imaged using an MRI scanner. The R2 (=1/T2) maps were extracted from several sequential T2-weighted MRI images. The dose values resulting from the polymer gel measurements at the reference points were compared with those from the common calculation method at the same points. In addition, the isodose curves resulting from gel dosimetry were compared with those from a brachytherapy treatment planning system (Flexiplan).
Results
The average of the dose values measured with the gel at the reference points was 62.75% higher than those calculated at the same points. Investigating the isodose curves revealed that the maximum distance to agreement (DTAmax) between the isodoses resulting from the gel and those obtained from the treatment planning system was less than 3 mm at different dose levels. Discussion and
Conclusion
Although the MAGICA gel indicates a higher absolute dose value than those calculated commonly, it can give the relative dose values accurately. Therefore, it can be recommended to be used for the assessment of dose distributions for the treatment of tissues as well as quality control of the treatment planning systems.
Language:
Persian
Published:
Iranian Journal of Medical Physics, Volume:7 Issue: 29, 2011
Page:
23
magiran.com/p837554  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!