Assessment of the Accuracy, Precision and Reproducibility of Apparent Diffusion Coefficient Measurements on a Clinical Magnetic Resonance Imaging System

Article Type:
Research/Original Article (بدون رتبه معتبر)
Abstract:
Purpose
The purpose of our work was to investigate delakis[1] quality control protocol for our MRI system on Tissue-equivalent diffusivity phantoms made of Nickel-doped agarose/sucrose gels.
Materials And Methods
we designed and manufactured a spherical phantom using these gels for T1, T2 and DW-MRI. Every compartments was filled with tissue equivalent relaxation and diffusivity gels. After that we assessed the quality control protocol for T1, T2 and DW-MRI on these gels and The magnetic resonance imaging of the phantom was performed on the 1.5 T clinical MRI system IMAM KHOMEINI hospital of TEHRAN (GE GENESIS SIGNA). Two parameters,Q and R, are used for the analysis of the quality control ADC values.
Results
T2 gel values, 84.804, 80.773, 86.57, 77.774, 77 (ms), were respectively obtained for BT(body tissue), P2(tumur), P2 in air, P2 in oil, P1(ischemi). Also, their corresponding T1 measurements were 1090.92, 1742.75, 1284.75, 1400, 1358.23 (ms) respectively. using this phantom DW- MRI experiments can be performed under very realistic conditions. The Q calculated for the p1 and p2 in different b-values (150, 400, 1000 mm s-2 ) is smaller than the Q for BT. This observation is likely to indicate that for b values 150, 400, 1000 mm s-2, the ADC measurement reproducibility is compromised by ADC deviations due to difference between effective and nominal b values over time. the R parameter used in our quality control protocol to quantitatively study the directional dependence is expected to be governed by the ADC fluctuation described by the equation of measurement of differences between nominal and effective b values.
Conclusion
. the quality control protocol presented in this study will be of value for monitoring the diffusion imaging performance of an MRI system in a clinically relevant way.
Language:
English
Published:
Frontiers in Biomedical Technologies, Volume:4 Issue: 3, Summer -Autman 2017
Pages:
49 to 58
magiran.com/p1831544  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!