A New Sonographic Phantom for Quality Control and Training Purposes

Message:
Abstract:
Background
Evaluation of the accuracy and performance of sonography units needs tissue-mimicking phantoms. These phantoms play an important role by simulating soft tissues, obviating the need to experiment on humans or animals..
Objectives
To present a simple sonographic phantom for quality control and training purposes..
Materials And Methods
The presented phantom consists of a two-part Plexiglas box. The larger part is filled with a mixture of ethanol (9.5 ± 0.25%) in distilled water and a solution of sodium nitrite (0.1 M) to prevent rusting. The second part is filled with a combination of 3.85% by wt. % agar, and 50 g/L of powdered graphite as the background material. In this study, chrome-plated electric guitar strings, 0.52 mm in diameter, were used. Several objects were considered as tissue-equivalent material, and their images were obtained at different times. Criteria for the selection of suitable objects comprised similarity between the obtained image and the corresponding tissues in the human body, minimal shrinkage, and change in brightness level at different times. In addition to quantitative analysis obtained from image processing, a blind qualitative study was done by a radiologist..
Results
Both results of quantitative analysis using MATLAB software and independent qualitative analysis showed that the commercial rubber and agar were appropriate as solid and cystic objects, respectively. Moreover, quantitative analysis done with MATLAB on images obtained from the phantom showed that the commercial rubber and agar had a 5% and 2% change in image pixel intensity (brightness) after 2 months, respectively..
Conclusions
The presented phantom not only has lower cost and complexity, which make it suitable for educational centers, but also is capable of providing good images of cystic and solid objects for quality control and training purposes. Furthermore, it confers reliable stability for at least 2 months, as was assessed in the present study..
Language:
English
Published:
Archives Of Cardiovascular Imaging, Volume:2 Issue: 2, May 2014
Page:
2
magiran.com/p1279792  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
دسترسی سراسری کاربران دانشگاه پیام نور!
اعضای هیئت علمی و دانشجویان دانشگاه پیام نور در سراسر کشور، در صورت ثبت نام با ایمیل دانشگاهی، تا پایان فروردین ماه 1403 به مقالات سایت دسترسی خواهند داشت!
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!