Performance Evaluation of Bone Mineral Densitometry techniques by a Novel Phantom

Message:
Abstract:
Purpose
Accurate performance assessment of bone mineral density (BMD) methods is crucial due to the fact that a high level of exact estimation of bone situation is needed for correct diagnosis. Variation of parameters like sensitivity and error ratio highly affects the densitometry results that may induce some level of uncertainty in diagnosis. So, designing an algorithm for correction is necessary to assure examiners about measurement results.
Methods
In this study several phantoms consisting of soft tissue- and bone-equivalent materials were devised to accurately test bone densitometry systems. To the best of our knowledge there is no unique phantom to be able to use for evaluation of both Dual Energy X-ray Absorptiometry (DEXA) and Quantitative Computed Tomography (QCT) in a wide range of density. The main motivation of this study was to design a reliable and easy to use phantom. A QCT quality control, quality assurance, and Plexiglas cylindrical phantoms as a spine phantom were designed and constructed to assess different bone densities. Four inserts in spine phantom with precisely wide range of K2HPO4 solutions were used for simulation of bone tissues and to determine the BMD systems characteristics. The designed phantoms were also used for performance assessment of BMD systems. We used a sinogram-based analytical CT simulator to model the complete chain of CT data acquisition for QCT method as well.
Results
In this research it is demonstrated that by decreasing of bone mineral densities an increasing trend in error ratio of measured densities and declining trend in methods sensitivities were observed in the both DEXA and QCT methods, that may cause some level of uncertainty in low densities. It has been shown that between the ranges of 20 and 100 mg/cc K2HPO4 concentrations, the error ratio in both DEXA and QCT techniques is more than 20%. Sensitivity values in incremental mineral contents ranges between 20-60 mg/cm3and 260-300 mg/cm3 reveal an upward trend between 0.93 and 1.45 for QCT and from 0.59 to 1.44 for DEXA, respectively.
Conclusion
A novel phantom was designed with capability of easily supporting wide range of densities and using in both DEXA and QCT techniques to measure and compare the sensitivity and error of systems. Our phantom showed excellent capability for accurate determination of BMD, particularly in low density bones. In this study it is demonstrated that the sensitivity and error ratio is affected by bone density that may cause uncertain results especially in low densities.
Language:
English
Published:
Frontiers in Biomedical Technologies, Volume:1 Issue: 4, Autumn 2014
Pages:
271 to 278
magiran.com/p1387427  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
دسترسی سراسری کاربران دانشگاه پیام نور!
اعضای هیئت علمی و دانشجویان دانشگاه پیام نور در سراسر کشور، در صورت ثبت نام با ایمیل دانشگاهی، تا پایان فروردین ماه 1403 به مقالات سایت دسترسی خواهند داشت!
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!