Evaluation of the Quality Control Program for Diagnostic Radiography and Fluoroscopy Devices in Syria during 2005-2013

Abstract:
Introduction
Extensive use of diagnostic radiology is the largest contributor to total population radiation doses. Thus, appropriate equipment and safe practice are necessary for good-quality images with optimal doses. This study aimed to perform quality control (QC) audit for radiography and fluoroscopy devices owned by private sector in Syria (2005-2013) to verify compliance of performance of X-ray machines with the regulatory requirements stipulated by the national regulatory body.
Materials And Methods
In this study, QC audit included 487 X-ray diagnostic machines, (363 radiography and 124 fluoroscopy devices), installed in 306 medical diagnostic radiology centers in 14 provinces in Syria. We employed an X-ray beam analyzer device (NERO model 8000, Victoreen, USA), which was tested and calibrated at the National Secondary Standard Dosimetry Laboratory traceable to the IAEA Network of Secondary Standard Dosimetry Laboratories. Standard QC tool kits were used to evaluate tube and generator of the X-ray machines, which constituted potential (kVp), timer accuracy, radiation output consistency, tube filtration, small and large focal spot sizes, X-ray beam collimation and alignment, as well as high- and low-resolution and entrance surface dose in fluoroscopy.
Results
According to our results, most of the assessed operating parameters were in compliance with the standards stipulated by the National Regulatory Authority. In cases of noncompliance for the assessed parameters, maximum value (28.77%) pertained to accuracy of kVp calibration for radiography units, while the lowest value (2.42%) belonged to entrance surface dose in fluoroscopy systems.
Conclusion
Effective QC program in diagnostic radiology leads to obtaining information regarding quality of radiology devices used for medical diagnosis and minimizing the doses received by patients and medical personnel. The findings of this QC program, as the main part of QA program, illustrated that most of the considered diagnostic X-ray devices had
acceptable performance and few of them need to be recalibrated for some parameters.
Language:
English
Published:
Iranian Journal of Medical Physics, Volume:14 Issue: 2, Spring 2017
Pages:
92 to 97
magiran.com/p1701796  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
دسترسی سراسری کاربران دانشگاه پیام نور!
اعضای هیئت علمی و دانشجویان دانشگاه پیام نور در سراسر کشور، در صورت ثبت نام با ایمیل دانشگاهی، تا پایان فروردین ماه 1403 به مقالات سایت دسترسی خواهند داشت!
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!