Effect of Flip Angle on the Correlation Between Signal Intensity and Different Concentrations of Iron Oxide Nanoparticles Using T1-Weighted Turbo-FLASH Inversion Recovery Sequence

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Background
Ultrasmall superparamagnetic iron oxide nanoparticles have been used as a blood pool contrast agent for magnetic resonance angiography and perfusion studies. Linear relationship between signal intensity (SI) and nanoparticle concentration is essential for perfusion measurement.

Objectives
The aim of this study was to investigate the effect of different flip angles on maximum SI and the linear relationship between SI and different concentrations of iron oxide nanoparticles using T1-weighted Turbo-FLASH (fast low angle shot) inversion recovery sequence to find the optimum flip angle for perfusion measurement.

Materials and Methods
This in vitro study was performed using carboxydextran coated iron oxide nanoparticles with 20 nm hydrodynamic size. Different concentrations of nanoparticles between 0 and 500 µmol Fe/L were prepared. MR imaging was performed using T1-weighted Turbo-FLASH inversion recovery sequence. Applied flip angles were 10-45º (interval of 5º). Then the maximum SI resulted by each concentration of nanoparticles was measured. Linear relationship between SI and nanoparticle concentration was evaluated regarding square correlations of 0.95 and 0.99. Coil non-uniformity was considered to obtain accurate SI of each image.

Results
The maximum SI was obtained at the highest applied flip angle (45°). The linear relationship between SI and nanoparticle concentration was seen up to 112.21 and 98.83 μmol Fe/L for the short (10°) and the long (45°) flip angles, respectively (R2 = 0.95). These values were reduced up to 48.54 and 42.73 μmol Fe/L for these flip angles with R2 of 0.99.

Conclusions
The maximum SI will be increased at higher flip angles with non-linear relationship between SI and nanoparticle concentration. The result shows that an increase in the flip angle leads to a decrease in the range of the linearity. The optimum flip angle which is suitable for perfusion measurement was obtained at 10º for our imaging parameters and sequence. The results of this study may be used in in vivo perfusion measurements.
Language:
English
Published:
Iranian Journal of Radiology, Volume:12 Issue: 2, Apr 2015
Page:
12
magiran.com/p1980057  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
دسترسی سراسری کاربران دانشگاه پیام نور!
اعضای هیئت علمی و دانشجویان دانشگاه پیام نور در سراسر کشور، در صورت ثبت نام با ایمیل دانشگاهی، تا پایان فروردین ماه 1403 به مقالات سایت دسترسی خواهند داشت!
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!