Impact of Iterative Reconstruction Algorithms on Image Quality and Radiation Dose in Computed Tomography Scan of Patients with Malignant Pancreatic Lesions

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Background

The objective of this study was to investigate the influence of iterative reconstruction (IR) algorithm on radiation dose and image quality of computed tomography (CT) scans of patients with malignant pancreatic lesions by designing a new protocol.

Methods

The pancreas CT was performed on 40 patients (23 males and 17 females) with a 160‑slice CT scan machine. The pancreatic parenchymal phase was performed in two stages: one with a usual dose of radiation and the other one after using a reduced dose of radiation. The images obtained with usual dose were reconstructed with Filtered Back Projection (FBP) method (Protocol A); and the images obtained with the reduced dose were reconstructed with both FBP (Protocol B) and IR method (Protocol C). The quality of images and radiation dose were compared among the three protocols.

Results

Image noise was significantly lower with Protocol C (10.80) than with Protocol A (14.98) and Protocol B (20.60) (P < 0.001). Signal‑to‑noise ratio and contrast‑to‑noise ratio were significantly higher with Protocol C than with Protocol A and Protocol B (P < 0.001). Protocol A and Protocol C were not significantly different in terms of image quality scores. Effective dose was reduced by approximately 48% in Protocol C compared with Protocol A (1.20 ± 0.53 mSv vs. 2.33 ± 0.86 mSv, P < 0.001).

Conclusion

Results of this study showed that applying the IR method compared to the FBP method can improve objective image quality, maintain subjective image quality, and reduce the radiation dose of the patients undergo pancreas CT.

Language:
English
Published:
Journal of Medical Signals and Sensors, Volume:12 Issue: 1, Jan-Mar 2022
Pages:
69 to 75
https://www.magiran.com/p2372238  
سامانه نویسندگان
  • Chaparian، Ali
    Corresponding Author (2)
    Chaparian, Ali
    Full Professor Medical Physics, Department of Medical Physics, Faculty of Medicine, Isfahan University Of Medical Sciences, اصفهان, Iran
اطلاعات نویسنده(گان) توسط ایشان ثبت و تکمیل شده‌است. برای مشاهده مشخصات و فهرست همه مطالب، صفحه رزومه را ببینید.
مقالات دیگری از این نویسنده (گان)