Purpose

The concept of Quality Control (QC) is considered a regular method to control, stabilize, and inspect the function of the diagnostic imaging system. The objective of implementing the QC program is to produce high-quality images by applying a minimum dose of radiation based on the As Low As Reasonably Achievable (ALARA) principle. Therefore, this study aimed to evaluate the status of radiation protection in diagnostic radiology wards of educational hospitals affiliated with Hamadan University of Medical Sciences.

Materials and Methods

In order to implement the QC programs, standard QC tests were performed for 11 devices at educational hospitals affiliated with Hamadan University of Medical Sciences. A Sweden QC kit called Pirranha was used to carry out the QC tests of X-ray devices, and the dosimetry of controlled areas. Also, the measurement of ambient dose in different places was performed by Graetz dosimeter made in Germany.

Results

Voltage Reproducibility, Exposure time reproducibility, tube outlet Linearity, and tube outlet reproducibility tests in all radiology departments which were in accordance with standard criteria were accepted; however, about 10% of the total filtration resulted in different centers needed to be corrected. In terms of radiation protection, 5% of the centers had problems related to warning signs, dimensions of radiology rooms were not standard at 7% of wards and also required protection was not sufficient at 9 percent. Moreover, there were problems with 12% of radiology centers in terms of dosimetry results and the efficiency of different parts of the radiology device.

Conclusion

QC programs performed by authorized companies are costly. But if these programs are done by qualified physicists in addition to reducing costs, we will see a significant increase in the accuracy and precision of the obtained results.

Language:
English
Published:
Frontiers in Biomedical Technologies, Volume:9 Issue: 3, Summer 2022
Pages:
199 to 205
magiran.com/p2450405  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!