Reducing scan time using time-of-flight protocol in clinical [18F]FDG-PET/CT imaging: A feasibility study

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Introduction

This study evaluates scan time reduction using time-of-flight (TOF) PET, when quantitative parameters including volumetric measures are considered.

Methods

32 patients were included in the study. The reconstruction parameters for TOF were 2 iterations, 18 and 24 subsets, and for non-TOF was 3 iterations and 18 subsets. A post smoothing filter with FWHM of 5.4 mm and 6.4 mm were used for TOF and 6.4 mm for non-TOF. TOF reconstruction was performed with 2, 2.5 and 3 min/bed position, and 3 min/bed position scan time was applied in non-TOF. Quantitative parameters such as coefficient of variation (COV), signal-to-noise ratio (SNR), lesion-to-background ratio (LBR), metabolic tumor volume (MTV) and total lesion glycolysis (TLG) were utilized. Standard uptake value (SUV) was also measured. Different segmentation thresholds were studied.

Results

Improvement in SNR for TOF relative to non-TOF was observed when utilizing 18 subsets, 5.4 mm filter size with 3 min scan time/bed position (P-value<0.0001), and for 18 subsets, 6.4 mm filter size when 2.5- and 3-min scan time/bed positions was applied (P-value≤0.02). Scan time reduction did not illustrate significant variation for the SUVs and lesion size. In all TOF protocols for both TLG and MTV, the measured values decreased with increasing segmentation thresholds, as expected, with significantly more impact for higher thresholds (70%, 75%). Meanwhile, higher values were observed for higher post smoothing filter in each specified threshold. With increasing of the threshold, ΔTLG was increased with more impact for higher post smoothing filter. ΔMTV were -10.10±11.09 (-1.35±8.59) and 0.68±17.51 (12.42±18.90) in 2 min/bed position with 5.4 (6.4 mm post smoothing filter) for threshold of 45% and 75% respectively.

Conclusion

Scan time reduction from 3 to 2 min can be obtained with TOF in comparison with non-TOF, especially when higher segmentation threshold values with higher subset number (24) and 6.4 mm filters are utilized.

Language:
English
Published:
Iranian Journal of Nuclear Medicine, Volume:31 Issue: 1, Winter-Spring 2023
Pages:
1 to 10
magiran.com/p2536253  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
دسترسی سراسری کاربران دانشگاه پیام نور!
اعضای هیئت علمی و دانشجویان دانشگاه پیام نور در سراسر کشور، در صورت ثبت نام با ایمیل دانشگاهی، تا پایان فروردین ماه 1403 به مقالات سایت دسترسی خواهند داشت!
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!