Safety and Precision of Two Different Flap-morphologies Created During Low Energy Femtosecond Laser-assisted LASIK

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Purpose

Currently, two major principles exist to create LASIK flaps: firstly, a strictly horizontal (2D) cut similar to the microkeratome-cut and secondly an angled cut with a “step-like” edge (3D). The strictly horizontal (2D) cut method can be performed using apparatus such as the low-energy FEMTO LDV Z8 laser and its predecessors which are specific to this type. Alternatively, the low-energy FEMTO LDV Z8 laser’s 3D flap design creates an interlocking flap-interface surface which potentially contributes toward flap stability. In addition, the FEMTO LDV Z8 offers flap-position adjustments after docking (before flap-creation). The current study analyzed precision, safety, efficacy, as well as patient self-reported pain and comfort levels after applying two different types of LASIK flap morphologies which were created with a low-energy, high-frequency femtosecond (fs) laser device.

Methods

A prospective, interventional, randomized, contralateral eye, single-center comparison study was conducted from November 2019 to March 2020 at the Hamburg vision clinic/ zentrumsehstärke, Hamburg, Germany. Eleven patients and 22 eyes received low-energy fs LASIK treatment for myopia or myopic astigmatism in both eyes. Before the treatment, the eyes were randomized (one eye was treated with the 2D, the other eye with the 3D method).

Results

The mean central flap thickness one month after surgery was 110.7 ± 1.6 μm (2D) and 111.2 ± 1.7 μm (3D); P = 0.365 (2D vs 3D). Flap thickness measured at 13 different points resulted in no statistically significant differences between any of the measurement points within/between both groups; demonstrating good planarity of the flap was achieved using both methods. Despite not being statistically significant, the surgeons recognized an increase in the presence of an opaque bubble layer in the 3D flap eyes during surgery and some patients reported higher, yet not statistically significant, pain scores in the 3D flap eyes during the first hours after the treatment. Overall, safety- and efficacy indices were 1.03 and 1.03, respectively.

Conclusion

In this prospective, randomized, contralateral eye study, the low-energy fs laser yielded predictable lamellar flap thicknesses and geometry at one-month followup. Based on these results, efficacy and safety of the corresponding laser application, that is, 2D vs 3D, are equivalent.

Language:
English
Published:
Journal of Ophthalmic and Vision Research, Volume:18 Issue: 1, Jan-Mar 2023
Pages:
3 to 14
magiran.com/p2546703  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!