Verification of the Accuracy of the Delivered Dose in Pelvic and Breast Cancer Radiotherapy by in-vivo Semi-Conductor Dosimetry

Message:
Abstract:
Introduction

Delivering maximum dose to tumor and minimum dose to normal tissues is the most important goal in radiotherapy. According to ICRU, the maximum acceptable uncertainty in the delivered dose compared to the prescribed dose should be lower than 5%, and this is because of the relationship between absorbed dose, tumor control and normal tissue damage. Absorbed dose accuracy is investigated by an in vivo dosimetry method. In this paper, we compared absorbed dose in the tumors of the breast and pelvic region against the calculated dose. The amount of deviations and the factors that cause this deviation in dose delivery to patients and some methods for decreasing them were evaluated.

Materials And Methods

The entrance and exit doses of 36 pelvic-region cancer patients and 38 breast cancer patients who were treated by cobalt-60 teletherapy were measured using p-type diodes. It should be noted that the transmission method was used to assess the dose at isocenter. Two ionization chambers (0.6 cc and 0.3 cc) were used for calibration and determination of the correction coefficients in water and slab phantoms. Deviations between calculated and measured doses of entrance, exit and midline points were calculated and the results were shown using histograms.

Results

The average and standard deviation for entrance, exit and midline points for pelvis cancer were assessed to be about 0.10%, -1.86% and -1.35% for mean deviation and 5.03%, 7.32% and 5.86% for standard deviation, respectively. The corresponding data for breast cancer were 0.78%, 5.29% and 3.59% for mean deviation and 5.97%, 10.23% and 9.86%, respectively. There was no significant difference between the calculated and measured doses (p > 0.1), except exit dose in breast cancer (p < 0.05). The temperature and angle of incidence correction factors were neglected due to their less than 1% deviations.Discussion and

Conclusions

Some error sources are patient setup error, patient motion and dose calculation algorithm error (due to ignoring inhomogeneity and patient curvature). As no significant deviations were found in midline dose, the method used has an acceptable accuracy. In vivo dosimetry can perform a basic role in the quality control of radiotherapy departments.

Language:
Persian
Published:
Iranian Journal of Medical Physics, Volume:6 Issue: 24, 2010
Pages:
38 to 45
magiran.com/p722485  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
دسترسی سراسری کاربران دانشگاه پیام نور!
اعضای هیئت علمی و دانشجویان دانشگاه پیام نور در سراسر کشور، در صورت ثبت نام با ایمیل دانشگاهی، تا پایان فروردین ماه 1403 به مقالات سایت دسترسی خواهند داشت!
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!