Calibration constant of a silver activation counter used with plasma focus devices

Message:
Abstract:

The silver activation counters are commonly used for pulsed-neutron yield measurements especially in plasma focus devices. The counter normally consists of a Geiger-Muller tube along with silver foils and polyethylene (as a moderator), which is calibrated against an Am-Be radioisotope neutron source. The neutrons, after being slowed-down in the polyethylene, activate the silver foils. By measuring the foil activity with a Geiger-Muller counter, the neutron yield is determined. In the present paper, the activation counter’s calibration constant calculation using the MCNP4C code is explained. The calculated calibration constant is in good agreement with the experimental results.

Language:
Persian
Published:
Iranian Journal of Physics Research (IJPR), Volume:9 Issue: 4, 2010
Page:
305
magiran.com/p727699  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
دسترسی سراسری کاربران دانشگاه پیام نور!
اعضای هیئت علمی و دانشجویان دانشگاه پیام نور در سراسر کشور، در صورت ثبت نام با ایمیل دانشگاهی، تا پایان فروردین ماه 1403 به مقالات سایت دسترسی خواهند داشت!
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!