Effect of Tissue Inhomogeneities on Dosimetric Accuracy of Conformal Radiotherapy of Prostate Cancer Using Monte Carlo Simulation

Abstract:
Background and
Purpose
The human body is composed of various inhomogenous tissues with a variety of physical and radiological properties. These inhomogeneities could change isodose distributions, increase the probability of geometric errors, and eventually lead to missing of the target irradiation or incorrect isodose coverage in conformal radiation therapy (CRT) due to the uncertainties resulted from the effect of such inhomogeneties on isodose curves. In this study, the effect of inhomogeneities on dosimetric accuracy of the CRT of prostate was investigated using Monte Carlo simulation method.
Materials And Methods
An anthropomorphic pelvis phantom was used in two modes representing a homogeneous and heterogeneous phantom. Three dimensional (3D) CRT planning composed of five fields were performed on the CT images of the phantoms using the CorPLAN treatment planning system (TPS) of a 2100C/D Varian linac. Delivery of 3DCRT was also simulated by EGSnrc Monte Carlo code for the same conditions used in the CorPLAN TPS. The dose distributions resulted from the MC simulations were compared with those of the CorPLAN TPS for both of the phantoms based on the "dose difference (DD) percentages" and "distance to agreement (DTA)" parameters.
Results
Findings indicated that the heterogeneity leads to an overdose estimation at the target location up to 2.8% and 4.4% for the 6 and 18 MV energies, respectively by the CorPLAN TPS compared with the MC simulation data. The average DD and DTA for the homogeneous phantom were 2.7%-3.4mm and 1.6%-2.3 mm while for the inhomogeneous phantom they were 5.5%-3.7 mm and 6.0%-2.5 mm at the 6 and 18 MV energies, respectively.
Conclusion
Our results indicates that the accuracy of the dosimetry parameters estimated by the CorPLAN TPS at 18 MV is more than that of 6 MV energy. But, the heterogeneity deteriorates more the accuracy level of such dosimetry parameters estimated by this commercial TPS at 18 MV energy.
Language:
Persian
Published:
Journal of Mazandaran University of Medical Sciences, Volume:26 Issue: 137, 2016
Pages:
137 to 149
magiran.com/p1552023  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!