Active Personal Dosimeter in a Nuclear Medicine Center in Yazd City, Iran

Message:
Abstract:
Introduction
Active personal dosimeters (APDs) are well accepted as useful and reliable instruments for individual dosimetry measurements. APDs have many advantages compared with passive dosimeters for individual external radiation dose assessments. In routine monitoring, occupational exposure is carried out for verification and demonstration of compliance with the regulatory dose limits. So, it is one of the most important tools in order to achieve or demonstrate the level of radiation protection.
Methods
Yazd province has only one private nuclear medicine (NM) center. In this center, two NM technologists exposed to radioactive patients during radiopharmacuticals preparation were monitored. NM technologists have to be close to the patient during radiopharmaceutical injection and patient positioning on the gamma camera table. An electronic personal dosimeter DKG-21 Ecotest made in Ukraine which records the ambient dose equivalent rate and equivalent dose was used to monitor the radiation exposure to the technologists and to record the accumulation dose in mSv throughout a working day. This study was accomplished between the time period of January to June 2011. The dosimeter is designed to measure individual equivalent dose Hp(10). The dose range of gamma radiation was 0.01 mSv to 1 Sv and the energy range 0.05 to 6 MeV which was suitable for NM procedures. The planar and tomography NM images were performed by the 2 technologists in the morning and afternoon shifts.
Results
The average monthly occupational dose of each technologist was approximately 0.6 mSv. Their annual doses were 6.6 and 8.8 mSv, respectively. They were lower than the maximum permissible dose of 20 mSv/y. Total number of NM procedures performed in this NM center during June 2010 to June 2011 was 3265.
Conclusion
The use of APD for monitoring the NM technologists is a useful tool to check compliance with regulatory dose limits and radiation protection principals.
Language:
English
Published:
Journal of Biomedical Physics & Engineering, Volume:1 Issue: 1, Jan-Feb 2011
Page:
28
magiran.com/p1005720  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!