The Quality Control of Intensity Modulated Radiation Therapy (IMRT) for ONCOR Siemens Linear

Message:
Abstract:
Introduction
Intensity Modulated Radiation Therapy (IMRT) has made a significant progress in radiation therapy centers in recent years. In this method, each radiation beam is divided into many subfields that create a field with a modulated intensity. Considering the complexity of this method, the quality control for IMRT is a topic of interest for researchers. This article is about the various steps of planning and quality control of Siemens linear accelerators for IMRT, using film dosimetry. This article in addition to review of the techniques, discusses the details of experiments and possible sources of errors which are not mentioned in the protocols and other references.
Materials And Methods
This project was carried out in Isfahan Milad hospital which has two Siemens ONCOR linear accelerators.Both accelerators are equipped with Multi-Leaf Collimators (MLC) which enables us to perform IMRT delivery in the step-and-shoot method. The quality control consists of various experiments related to the sections of radiation therapy. In these experiments, the accuracy of some components such as treatment planning system, imaging device (CT), MLC, control system of accelerator, and stability of the output are evaluated. The dose verification is performed using film dosimetry method. The films were KODAK-EDR2, which were calibrated before the experiments. One of the important steps is the comparison of the calculated dose with planning system and the measured dose in experiments.
Results
The results of the experiments in various steps have been acceptable according to the standard protocols. The calibration of MLC and evaluation of the leakage through the leaves of MLC was performed by using the film dosimetry and visual check. In comparison with calculated and measured dose, more that 80% of the points have to be in agreement within 3% of the value. In our experiments, between 85 and 90% of the points had such an agreement with IMRT delivery.
Conclusion
The EDR2 films are suitable for quality control of IMRT. According to complexity of the quality control for IMRT, the physicists of each center have to develop specific guidelines according to their equipments and limitations. An accurate treatment planning system with capability of inverse planning is an essential need for IMRT. The result of the planning system has to be compared with experiments in various situations.
Language:
English
Published:
Iranian Journal of Medical Physics, Volume:9 Issue: 2, Spring 2012
Page:
111
magiran.com/p1041799  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!