Purpose
To investigate the effect of simultaneously performed fluorescein angiography (FA) on full-field electroretinography (ffERG) parameters.
Method
Scotopic and photopic ffERG were performed immediately and 60 minutes after conventional FA in patients with retinal photoreceptor disorders; a-and b-wave amplitudes were compared between recordings obtained at the two time intervals in each patient.
Results
Ten eyes of five (3 male and 2 female) patients with mean age of 19.6±3.8 (range, 15-25) years were studied. Intravenous fluorescein administration caused an immediate reduction in ERG waves which was most prominent in rod and maximal combined responses. Mean a-wave amplitude in maximal combined response, rod response and cone response ERGs was 46.0±18.8, 8.0±7.0 and 5.1±2.0? v immediately after FA which was increased to 79.0±30.0, 21.5± 22.5 and 6.5±2.4? v 60 minutes afterwards, respectively (P)
Language:
English
Published:
Journal of Ophthalmic and Vision Research, Volume:7 Issue: 4, Oct-Dec 2012
Pages:
300 to 304
magiran.com/p1075638  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!