The Dosimetric Effects of Different Multileaf Collimator Widths on Physical Dose Distributions

Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Introduction
Geometric changes in the multileaf collimator (MLC) led to dosimetric considerations in intensity-modulated radiation therapy (IMRT) due to the number and size of the pixels in the intensity map, which are determined by the MLC leaf width. In this study, we evaluated the dosimetric effects of different MLC widths on physical dose distributions for IMRT plans.
Materials And Methods
Forty-two IMRT plans based on different MLC devices were generated and analyzed to study the effect of MLC width on plan quality.
Results
Improvements in IMRT plan quality using 0.4 cm leaf width in comparison with 1 cm leaf width were evaluated. The 0.4 cm leaf-based plans resulted in significantly higher Dmean, D98%, D95% , D5%, and V95 (58.86 Gy, 95.11%, 96.57%, 104%, and 97.92%, respectively) compared to the 1 cm leaf plans (58.66 Gy, 92.56%, 94.56%, 104.14%, and 95.72%, respectively). Conformation number (CN) for planning target volume in 0.4 cm leaf plans was significantly higher than the 1 cm leaf plans (0.74 vs. 0.67; P
Conclusion
No significant clinical differences were observed between the two plans for a serially functioning tissue, while the differences in mean doses were statistically significant for parallel functioning normal tissues.
Language:
English
Published:
Iranian Journal of Medical Physics, Volume:15 Issue: 1, winter 2018
Pages:
12 to 18