Repeatability of curvature measurements in central and paracentral corneal areas of keratoconus patients using Orbscan and Pentacam

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Purpose

To determine the repeatability of curvature measurements in 5 corneal rings (1–5 mm from the corneal center) in keratoconus (KCN) patients using the Orbscan and Pentacam and to compare the values of these devices.

Methods

Forty-eight patients with a definite diagnosis of KCN were included in the study. Patients with any corneal scar or active disease or a history of ocular surgery were excluded from the study. The right eye of the patients was studied three times with the Orbscan and Pentacam. The repeatability of the curvatures of 5 corneal rings (1–5 mm from the corneal center) was evaluated using the Orbscan and Pentacam, and the agreement of their values was analyzed.

Results

The intraclass correlation coefficient (ICC) of three measurements was at least 0.94 (P < 0.0001) for the Orbscan and at least 0.88 (P < 0.0001) for the Pentacam in all corneal rings. According to the grade of KCN, the Orbscan had a low ICC in the 2 mm ring in grades 2 and 3 (ICC = 0.750 and 0.298, respectively). Repeated measures ANOVA showed no significant difference between the repeated measurements of the Orbscan and Pentacam in all corneal rings. The paired t-test showed a significant difference in curvature measurements in all rings except for the 5-mm ring between the two devices (P < 0.0001). The Bland-Altman plot showed a week agreement between these two devices in 1–4 mm corneal rings in curvatures more than 45 D.

Conclusions

According to the results of this study, keratometry readings are highly repeatable in Pentacam and Orbscan devices in all corneal rings. Despite the high correlation between curvature measurements of the Orbscan and Pentacam, there was a significant statistical and clinical difference between the results of two devices in all corneal rings (except the 5-mm ring), and the curvature measurements of the Pentacam were steeper than Orbscan measurements.

Language:
English
Published:
Journal of Current Ophthalmology, Volume:31 Issue: 4, Dec 2019
Pages:
382 to 386
magiran.com/p2057044  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!