A Study of the Impact of TRS-483 Recommended Correction Factors for Dosimetry of Small Fields in Flattening Filter Free Beams used in TrueBeam Linear Accelerators

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Introduction

To study the effect of the International Atomic Energy Agency (IAEA) TRS-483 recommended beam quality correction factor in reference dosimetry and to examine the recommended field output correction factor for relative dosimetry of 6-MV flattening filter free (FFF) small fields, used in a Varian TrueBeam linear accelerator (LINAC).

Material and Methods

The beam quality and field output correction for 6-MV FFF beams were adopted from the TRS-483 protocol. Monte Carlo (MC) simulation of the output factor was performed using the PENELOPE-based PRIMO software and compared with the TRS-483 corrected output factors. Two analytical anisotropic algorithm (AAA) models in the EclipseTM treatment planning system (TPS) were created; one with an output factor taken as the ratio of meter readings and one with an output factor obtained by multiplying the TRS-483 correction factor by the ratio of meter readings. Besides, box field and dynamic conformal arc (DCA) plans were created for both AAA models for verification and validation. The patient-specific quality assurances (QA) for ten different targets were performed, and deviations between the measured and TPS-calculated point doses in both models were examined.

Results

Separate beam quality correction factors for FFF beams in the TRS-483 protocol only resulted in an improvement of 0.1% in reference dosimetry. The TRS-483 corrected output factor was in a better agreement with the MC-calculated output factor. For a patient-specific QA of DCA plans, the output factor-corrected AAA dose calculation algorithm showed a better agreement between the measured and simulated doses. Also, there was a smaller deviation (1.2%) for the smallest target of 0.23 cc (8 mm equivalent sphere diameter) used in this study.

Conclusion

The field output factors for the LINAC small beams can be improved by incorporating the TRS-483 correction factors. However, the extent of improvement that can be expected depends on the source model of the calculation algorithm and how these well-generalized corrections are suitable for user beams and detectors.

Language:
English
Published:
Iranian Journal of Medical Physics, Volume:18 Issue: 3, May-Jun 2021
Pages:
178 to 186
magiran.com/p2266640  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!