Low‑dose Cone‑Beam Computed Tomography Reconstruction through a fast Three‑Dimensional Compressed Sensing Method Based on the Three‑Dimensional Pseudo‑polar Fourier Transform

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Background

Reconstruction of high quality two dimensional images from fan beam computed tomography (CT) with a limited number of projections is already feasible through Fourier based iterative reconstruction method. However, this article is focused on a more complicated reconstruction of three dimensional (3D) images in a sparse view cone beam computed tomography (CBCT) by utilizing Compressive Sensing (CS) based on 3D pseudo polar Fourier transform (PPFT).

Method

In comparison with the prevalent Cartesian grid, PPFT re gridding is potent to remove rebinning and interpolation errors. Furthermore, using PPFT based radon transform as the measurement matrix, reduced the computational complexity.

Results

In order to show the computational efficiency of the proposed method, we compare it with an algebraic reconstruction technique and a CS type algorithm. We observed convergence in <20 iterations in our algorithm while others would need at least 50 iterations for reconstructing a qualified phantom image. Furthermore, using a fast composite splitting algorithm solver in each iteration makes it a fast CBCT reconstruction algorithm. The algorithm will minimize a linear combination of three terms corresponding to a least square data fitting, Hessian (HS) Penalty and l1 norm wavelet regularization. We named it PP‑based compressed sensing‑HS‑W. In the reconstruction range of 120 projections around the 360° rotation, the image quality is visually similar to reconstructed images by Feldkamp‑Davis‑Kress algorithm using 720 projections. This represents a high dose reduction.

Conclusion

The main achievements of this work are to reduce the radiation dose without degrading the image quality. Its ability in removing the staircase effect, preserving edges and regions with smooth intensity transition, and producing high‑resolution, low‑noise reconstruction results in low‑dose level are also shown.

Language:
English
Published:
Journal of Medical Signals and Sensors, Volume:12 Issue: 1, Jan-Mar 2022
Pages:
8 to 24
magiran.com/p2372252  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!