Tantalum Versus Titanium Acetabular Cups in Primary Total Hip Arthroplasty: Current Concept and a Review of the Current Literature

Message:
Article Type:
Review Article (دارای رتبه معتبر)
Abstract:
Background

Primary total hip arthroplasty (THA) is becoming an increasingly popular and efficacious medical procedure. There have been a number of studies evaluating tantalum acetabular cups compared with the conventional titanium acetabular cups for use in total hip arthroplasties. We conducted a systematic review and summarize clinical studies comparing tantalum acetabular cups with the conventional titanium acetabular cups for use in primary total hip arthroplasties.

Methods

A literature search was performed to find all relevant clinical studies until March 2020, which then underwent a further selection criteria. The inclusion criteria was set as follows: Reporting on human patients undergoing primary total hip arthroplasty; Direct comparison between tantalum acetabular cups with conventional titanium acetabular cups for use in primary total hip arthroplasty; Radiological evaluation (cup migration, osteointegration); Clinical (functional scores, need for subsequent revision, patient-reported outcomes; Post-operative complications; Reporting findings in the English Language. After a thorough search a total of six studies were included in the review. The primary outcome measures were clinical outcomes, implant migration, change in bone mineral density and rate of revision and infection. 

Results

Tantalum is superior to titanium with regards to fewer radiolucencies, 100% survivorship at 12 years postoperatively, improved long-term implant osteointegration and survivorship as well as decreasing osteolysis and mechanical loosening. There has been no significant difference in radioisometric analysis, bone mineral density or Harris Hip Score. Revision and infection rates were found to be significantly lower in tantalum group at 10 years from pooled data of national joint registry (England and Wales), while it was found to be higher in the same at 9 years from pooled data of Swedish and Australian registry although this is not statistically significant.

Conclusion

The use of tantalum should be reserved for cases of high risk of failure or mechanical loosening, where failure of a contralateral joint occurred. The use of Tantalum carries lower risk of failure and infection. Further studies with longer follow-up would be useful in drawing further conclusions. Level of evidence: II

Language:
English
Published:
Archives of Bone and Joint Surgery, Volume:10 Issue: 5, May 2022
Pages:
385 to 394
magiran.com/p2426849  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!