Relationship between the Patients’ Setup Errors with Dosimetric and Radiobiologic Parameters in Whole Breast Radiotherapy

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Purpose

This study aimed to investigate the effect of the patients’ setup errors on dosimetric and radiobiologic parameters for left-sided Whole-Breast Irradiation (WBI) in three different radiotherapy techniques, including Intensity-Modulated Radiation Therapy (IMRT), Field-In-Field (FIF), and Conventional Wedge (CW).

Materials and Methods

Computed Tomography (CT) images of 10 female patients with early-stage left-sided breast cancer were used to simulate different radiotherapy techniques (IMRT, FIF, and CW). The dosimetric parameters; Conformity Index (CI), Homogeneity Index (HI), the dose received by at least 95% (D95%) of Planning Tumor Volume (PTV), the volume of lung and heart that respectively received at least 20% (V20%) and 40% (V40%) of the prescribed dose, as well as, the radiobiologic parameters, including Tumor Control Probability (TCP) and Normal Tissue Complication Probability (NTCP) were assessed for setup errors in patients. The setup errors were assessed by shifting the isocenters and gantry angles of the treatment plans.

Results

The D95% of the PTV for an isocenter misplacement plan in the posterior direction decreased by 66.99 (IMRT), 71.86 (CW), and 68.25% (FIF). The TCP of the PTV was reduced by 26.66, 39.16, and 36.97% for IMRT, CW, and FIF techniques, respectively. Increasing gantry angle by a ±10 degree caused a 43%, 41%, and 41% decrease in the D95% of IMRT, FIF, and CW techniques, respectively. The TCP values decreased about 18% in all three techniques with a ±10 degree gantry angle shift; however, the NTCP values of the heart and lungs increased for all three methods. The CI and HI values had significantly more changes with increasing setup errors in the IMRT than in the two techniques.

Conclusion

The radiobiologic parameters in IMRT were less sensitive to setup errors compared to FIF and CW techniques. The radiobiological parameters can help estimate the setup errors along with physical parameters during breast radiotherapy.

Language:
English
Published:
Frontiers in Biomedical Technologies, Volume:10 Issue: 3, Summer 2023
Pages:
268 to 276
magiran.com/p2606677  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!