Fabrication of Free Air Well Type Ionization Chamber and Calculational Assessment and Measurement of Its Operational Characteristics

Message:
Abstract:
Introduction
Well type ionization chamber is a measuring device which is used to determine the activity of brachytherapy sources. The chamber has a cylindrical volume in which a cylindrical tube is mounted in the middle of the chamber. For the measurements, the brachytherapy sources are transferred to the middle of the tube.
Materials And Methods
For designing the well type chamber, the measurement principals of well type chambers were considered and MCNP-4C code as a calculation tool was used. The designed chamber was simulated and the response of the chamber was evaluated. In this investigation, the chamber operational parameters such as operating voltage, leakage current, reproducibility, reference measuring point, recombination and polarization factors as well as response stability for 137Cs, 57Co and 241Am sources were studied.
Results
The chamber leakage currents at the operating voltage in comparison to the chamber response for the measurement of the above mentioned sources were negligible. The responses of the fabricated chamber for these sources are reproducible and its reference measurement position for these sources was obtained at 6 cm from the bottom of the chamber. The recombination factor for the well type chamber was negligible and the polarization factor is close to 1. Therefore, these two factors were not considered in the measurements. The reproducibility of the measurements in different intervals shows the stability of the chamber response for each source. Also the results of the chamber current measurement in term of source strength were compared to the response of the simulated chamber for different source positions and energy ranges of the used sources. Discussion and
Conclusion
The results show that the measurement of the reference positions for each source in the simulated and fabricated chamber is quite in a good agreement. Regarding the reliable operational properties of the fabricated chamber, this chamber can be used to determine the activities of brachytherapy sources in which the energy ranges are greater than 50 KeV.
Language:
Persian
Published:
Iranian Journal of Medical Physics, Volume:4 Issue: 16, 2007
Pages:
1 to 10
magiran.com/p580501  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!