Design and Fabrication of the Mechanical Part of a Prototype Multileaf Collimator System
Author(s):
Abstract:
Introduction
Multileaf collimator (MLC) is state of the art among the radiation field shaping systems for conformal radiotherapy techniques. Some advantages of this system are as follow: reduction of treatment time, higher precision in conforming the radiation field to the projection of the target volume and lowering the chance of errors made by technologists. The purpose of this research was to design and construct the mechanical section of a prototype MLC. Material And Methods
The MLC consists of a base plate, 52 leaves, 2 leaf carriages, 2 nuts and 52 triangular thread screws. The leaves are arranged in two opposing banks each containing 26 leaves. The carriages are used to extend the movement of the leaves. The components of the MLC are fabricated by machining processes. The MLC is designed to be mounted on the conventional linear accelerators as an add-on accessory. The leaves were made of aluminum. The dimensions of the leaves are 5 x 200 x 50 mm. Result
The collimator system weighs approximately 37kg excluding the leaves. The projected width of each leaf on the isocenter of a Varian linear accelerator is 10 mm. The movement of the leaves is accomplished by using the triangular thread screws. The screws required to have a low axial backlash which in turn results in lower precision positioning of the leaves. The positioning accuracy of each leaf is approximately 1.4 mm. Discussion and Conclusion
Mashhad MLC system, in it’s present form and design weighs 37 kg. but if Aluminum blades are replaced by tungsten ones then the over all weights of MLC may exceed the weight limit a linac may hold. Therefore futher modifications are required to reduce the MLC total weight. It is desirable to use ball screws and roller cages instead of triangular thread screws to achive a higher positioning pereision of the blades.Keywords:
Language:
Persian
Published:
Iranian Journal of Medical Physics, Volume:4 Issue: 16, 2007
Pages:
11 to 18
https://www.magiran.com/p580502
سامانه نویسندگان
مقالات دیگری از این نویسنده (گان)
-
Anti-Apoptosis Effects of Green Tea in Diagnostic and Therapeutic Doses of Ionizing Radiation
Najmeh Anbiaee, Mohammadtaghi Bahreyni Toossi, Hosein Azimian, Zoha Sahebnasagh, Sepideh Abdollahi Dehkordi, Majid Kianmehr, Samaneh Soudmand Salarabadi, Maryam Najafi Amiri *
Iranian Journal of Medical Physics, Jul-Aug 2024 -
Study of the Radio-Protective Effects of Henna Mixed With Vinegar on the Skin of Irradiated Rats
Mojgan Hasanzade, Mohammadtaghi Bahreyni Toossi, Masoumeh Gharib, Shirin Taraz Jamshidi, Hamid Gholamhosseinian, Shokouhozaman Soleymanifard *
Iranian Journal of Medical Physics, Jul-Aug 2024