Repeatability and Agreement of Anterior Corneal Curvature Measurement susing Orbscan and Eyesys Instruments in Keratoconic Eyes

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Purpose
The anterior corneal curvature has important role in diagnosis and management of keratoconus. The purpose of the present study is to determine the repeatability of anterior curvature of 5 corneal rings (1-5 mm from the corneal center) in keratoconus patients obtained by the Orbscan and Eyesys instruments and to determine the level of agreement of curvature measurements between two instruments.
Methods
In this prospective study, anterior curvatures of 5 corneal rings were measured using Orbscan and Eyesys in 48 right eyes of 48 keratoconic patients. The mean age of patients was 31.6±4.7 and the grades of keratoconus of patients according to the maximum keratometry were mild (≤50 D), moderate (50-55D) and severe (≥55D). Statistical analysis was performed using Intraclass Correlation Coefficient and Repeated measures ANOVA for repeatability of each device. Pearson’s correlation coefficient, paired t test and Bland-Altman plot were also used for comparison of two techniques.
Results
In this study, the ICC showed high repeatability for measurements of each devices (ICC Eyesys=0.99, ICC Orbscan≥0.95, p<0.0001) in all corneal rings. According to repeated measures ANOVA, there was no significant difference between the repeated measurements of each device in all corneal rings. The Pearson’s correlation analysis showed high correlation between the measurements of Orbscan and those of Eyesys (r≥0.84, p<0.0001) in all rings. According to the paired t test, there was a significant difference in curvature measurements of Eyesys and those of Orbscan in all corneal rings (p<0.001). Also, the Bland-Altman plots showed low agreement between curvature measurements of two devices in all corneal rings.
Conclusion
The present study showed that the anterior curvature measurements obtained by each of Orbscan and Eyesys instruments were repeatable in all corneal rings. Anterior curvature measurements of 5corneal rings obtained by Eyesys were steeper than those obtained by Orbscan and the two devices could not be replaced for anterior corneal curvature measurement in keratoconic eyes.
Language:
Persian
Published:
Journal of Paramedical Science and Rehabilitation, Volume:8 Issue: 3, 2019
Pages:
56 to 62
magiran.com/p2047415  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!