A Simulation Study With Electronic Dosimeter to Estimate Patient and Personnel Dose in Orthopedic Surgery

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Background

The use of mini C-arm is required in many orthopedic surgeries as an image-guided tool. Studies have shown an unnecessarily high dose to patients if exposures are not properly optimized. Also, scatter radiation to personnel may increase the risk of cancer if an appropriate protective device is not used. 

Objectives

This study aims to determine the patient’s dose and scatter radiation to the surgeon, anesthetist, and scrub nurse. Also, a comparison will be made with other studies on this topic. 

Methods

A phantom was designed to simulate a patient on the operating table to produce scatter radiation. In the same vein, a mobile mini C-arm unit was positioned with the x-ray tube beneath the head and the image receptor above the operating table. Measurements were made with a measuring tape from the central axis to the position of the surgeon, anesthetist, and scrub nurse. The Entrance Surface Dose (ESD) was determined by placing the electronic dosimeter at the surface of the phantom to estimate the patient dose. Similarly, each personnel dose/dose rate from the central axis was determined using a similar detector. 

Results

The total average time for most orthopedic surgeries in the studied facility was 3.3 minutes. The estimated ESD to the patient was 25.03 µSv and the scatter radiation per patient reaching the simulated anesthetist, scrub nurse, and surgeon was, 3.75, 3.59, and 7.72 µSv, respectively. The estimated dose values per year to anesthetist, scrub nurse, and surgeon were 390, 373.36, and 802 µSv, respectively. 

Conclusion

The personnel recommended limit dose rate was <20000 µSv/y. The technical factor used and total exposure time from this study could have affected the radiation dose. This study showed that personnel was safe even without the use of lead apron. Nevertheless, the use of an appropriate protective device should be encouraged to ensure safety.

Language:
English
Published:
Journal of Research in Orthopedic Science, Volume:7 Issue: 2, May 2020
Pages:
61 to 66
magiran.com/p2140290  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!