Measuring the patient's skin dose in common radiographs in Loghman Hakim Hospital

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Background and purpose

Most exposure of people to artificial sources through diagnostic radiographs is due to the increase of these tests in the community and the inevitability of these tests. Optimizing the dose of skin inlet can play an important role in reducing radiation exposure, which can be achieved by monitoring the inlet dose as a key parameter.The purpose of this study was to evaluate the rate of skin input dose in routine radiographic tests at Loghman Hakim Hospital in Tehran to compare the doses recommended by the International Commission for Protection against Radiation (ICRP) as well as, if possible, to provide simple guidelines for dose reduction and optimization. Patients with preserved image quality.

Materials and Methods

This descriptive study was performed in Loghman Hakim public hospital in Tehran with 235 patients 100 females and 135 males with a weight range of ± 10 kg. In this study, skin doses of patients were measured using four diagnostic radiographs of IAEA standards, in 12 common radiographic views using thermoluminescence chips and Geiger Müller dosimeter.

Results

The doses of skin for each radiographic method in milligrays were as follows: Chest (posterior-anterior) 0.34, Chest (1.85), Abdominal (anterior-posterior) 3.31 , Hip (anterior-posterior) 4.79, cervical (anterior-posterior) 1.95, cervical vertebra (1.24), dorsal (anterior-posterior) 2.38, dorsal vertebrae (profile) 4.68), lumbar vertebrae 3.95, lumbar vertebrae 9.53, skull (antero-posterior) 29.2 and skull 1.89.

Conclusion

This study shows that the rate of skin inlet dose in the twelve investigated facets was in accordance with the recommended doses and was in line with the standards provided by the International Commission for Protection against ICRP and also optimized the radiation parameters. And standardizing radiographic conditions by preserving the image quality will reduce the dose received by patients. In this regard, dosimetry at regular intervals and periodic and regular quality control tests of radiographic devices can be optimized, monitored, reduced dose. And protecting patients as well as assessing potential harm The object of radiation will be effective.

Language:
Persian
Published:
Journal of Research In Medical Sciences, Volume:44 Issue: 3, 2020
Pages:
460 to 465
magiran.com/p2167397  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!