Comparison of Bioimpedance Versus Pulse Contour Analysis for Intraoperative Cardiac Index Monitoring in Patients Undergoing Kidney Transplantation

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Background

Cardiac index (CI; cardiac output indexed to body surface area) is routinely measured during kidney transplant surgery. Bioimpedance cardiometry is a transthoracic impedance as the non-invasive alternative for hemodynamic monitoring, using semi-invasive uncalibrated pulse wave or contour (UPC) analysis.

Objectives

We performed a cross-sectional observational study on 50 kidney transplant patients to compare the CI measurement agreement, concordance rate, and trending ability between bioimpedance and UPC analysis.

Methods

For each patient, CI was measured by bioimpedance analysis (ICONTM) and UPC analysis (EV1000TM) devices at three time points: after induction, during incision, and at reperfusion. The device measurement accuracy was assessed by the bias value, limit of agreement (LoA), and percentage error (PE) using Bland-Altman analyses. Trending ability was assessed by angular bias and polar concordance through four-quadrant and polar plot analyses.

Results

From each time point and pooled measurement, the correlation coefficients were 0.267, 0.327, 0.321, and 0.348. Bland-Altman analyses showed mean bias values of 1.18, 1.06, 1.48, and 1.30, LoA of -1.35 to 3.72, -1.39 to 3.51, -1.07 to 4.04, and -1.17 to 3.78, and PE of 82.21, 78.50, 68.74, and 74.58%, respectively. Polar plot analyses revealed angular bias values of -10.37º, -15.01º, -18.68º, and -12.62º, with radial LoA of 89.79º, 85.86º, 83.38º, and 87.82º, respectively. The four-quadrant plot concordance rates were 70.77, 67.35, 65.90, and 69.79%. These analyses showed poor agreement, weak concordance, and low trending ability of bioimpedance cardiometry to UPC analysis.

Conclusions

Bioimpedance and UPC analysis for CI measurements were not interchangeable in patients undergoing kidney transplant surgery. Cardiac index monitoring using bioimpedance cardiometry during kidney transplantation should be interpreted cautiously because it showed poor reliability due to low accuracy, precision, and trending ability for CI measurement.

Language:
English
Published:
Anesthesiology and Pain Medicine, Volume:11 Issue: 5, Oct 2021
Page:
3
magiran.com/p2352221  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!