Design of a computer software for calculation of required barrier against radiation at the diagnostic x-ray units

Abstract:
Background and
Purpose
Instalation of protective barrier against diagnostic x-ray is generally done based on the recommendations of NCRP49. There are analytic methods for designing protective barriers howerer, they lack sufficient efficiency and considering the NCRP49 reports, designing mechanical protective barrier in order to protect the initial x-ray radiation and absorption of the ray quality of such radiation is different. Therefore, the protective barrier for each radiation is measured separately. In this study, a computer software was designed to calculate the needed barrier with high accuracy.
Materials And Methods
Calculation of required protective barrier particularly when two or more generators are in use at diagnostic x-ray units and or installed diagnostic equipments do not have proper room space and the limitations for other clanges in parameters which are time- consuming and impossible to be manually calculated. For proper determination of thichness of the protective barrier, relevant information about curves of radiation weakness, dose limit etc should be entered. This program was done in windows and designed in such a way that the operator works easily, flexibility of the program is acceptable and its accuracy and sensitivity is high.
Results
Results of this program indicate that, in most cases, in x-ray units required protective barrier was not used. Meanwhile sometimes shielding is more than what required which lacks technical standards and cost effectiveness. When the application index is contrasting zero, thichness of NCRP49 calculation is about 20% less than the calculated rate done by the method of this study. When the applied index is equal to zero (that is the only situation where the second barrier is considered), thickness of requined barrier is about 15% less than the lead barrier and concrete barrier calculated in this project is 8% less than that calculated by McGuire method.
Conclusion
In this study proper determination of required protective thickness for weakening of radiation to the permitable level is proposed. The multi radiation sources in a single room are considered and non garanteed radiation of NCRP49 hypothesis are removed. Difference between the the oritical and calculated rates this of method is X2=10-5 which indicates accuracy and high efficiency of this software.
Language:
Persian
Published:
Journal of Mazandaran University of Medical Sciences, Volume:15 Issue: 47, 2005
Page:
90
magiran.com/p245750  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!