The effect of steam sterilization without preparation steps on the accuracy of friction-style MTLDs

Message:
Abstract:
Background and Aim
Inadequate torque application during screw tightening in dental Implants is one of the responsible factors of screw loosening and fracture and others technical complications. Mechanical torque limiting Devices (MTLDs) are necessary tools to control a peak torque and achieving a target values. Accuracy of these devices, particularly in friction–style type is reported with high variability. Due to vague and controversial aspect of steam sterilization on the accuracy of these devices their preparation steps (dismantling of the components, lubrication, broken position of the handle) before sterilization, this study aimed to evaluate the effect of steam sterilization without preparation steps on the accuracy of friction-style MTLDs.
Materials and Methods
Fifteen new friction-style MTLDs from three different manufacturers, five for each group, were purchased. All peak torque measurement, considering their target torque, were measured ten times consecutively before and after 100 cycles of steam sterilization using Tohnichi torque gauge. For steam sterilization procedure, 100 cycles of autoclave without preparation steps before each cycle were performed. Comparison of error values and range of difference between the measured torque value and the target torque values before/after sterilization were performed with a One- way Repeated- measures ANOVA to significant level of P<0.05, considering type of devices as a between subject comparison. For pair wise comparison Bonferroni post hoc test was used.
Results
After steam sterilization without preparation steps, difference between target torque and peak values increased in all of the 3 groups studied (P<0.05). Peak torque values showed more than 10% difference from the target torque (suitable clinically accepted torque) in some measurements. Range of error was -1 to 4 Ncm in Dr Idhe group, -1 to 3 Ncm in AstraTech group and -3.5 to 1 Ncm in Biohorizon group.
Conclusion
Within the limitation of this study maximum error values were 12% in Astra Tech MTLDs, 13.3% in Dr Idhe MTLDs and 11.7% in Biohorizon MTLDs. Greater variability was seen for Biohorizon and Dr Idhe groups than for Astra Tech group. CLINICAL IMPLICATION: Sterilization without preparation steps affects the accuracy of friction-style mechanical torque limiting devices. Combined effects of sterilization and aging needs to be determined.
Language:
Persian
Published:
Journal of Dental School, Volume:29 Issue: 5, 2012
Page:
306
magiran.com/p988785  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!