Evaluation of Absorbed Dose of Critical Organ in Rando Phantom under Head, Abdomen and Pelvis Spiral CT Scan by Thermo Luminescent Dosimetery (TLD)

Message:
Abstract:
Background and Objective
Computed tomography (CT) represents 11% of all diagnostic radiology procedures but it contributes to almost 67% of the total effective dose to the human population. In head and neck CT which consist of 1/3 of total CT scans, other critical organs such as lenses and thyroid are in the radiation field. Also in the abdomen and pelvis scan, irradiation of ovaries is unavoidable. Because of high sensitivity of these organs, the probability of abnormality and cancer in these organs has increased. Therefore the dose assessment in these organs is very important. The aim of this study is to estimate the absorbed dose in critical organ of patient undergoing common head, neck, abdomen and pelvic spiral CT scan.
Materials and Methods
In this study, Lithium fluoride thermo luminescent dosimeters (TLD-100, Harshaw) were used to determine the absorbed dose of critical organ of tissue equivalent rando phantom (Alderson research industries, Inc, Stanford, Conn, U.S.A). The phantom was sectional in design and manufactured with a 2.5 cm slab thickness. Each section contained some holes that allowed accommodation of TLDs. At least two crystals were placed in each hole. The average value of the TLD readings was taken as the organ dose. Readouts were obtained on a Harshaw 4500 reader (Harshaw, Ohio, USA). For calibration, the annealed dosimeters were exposed to an X-ray beam resulting from 120 kVp tube voltage and calibration curve was plotted.
Results
Result of this study showed during head CT scan the maximum absorbed dose belongs to occipital bones skin. Which were about 11.45 mGy and the minimum absorbed dose belong to thyroid gland which was 0.5 mGy. During abdomen & pelvic spiral CT, the maximum absorbed dose of abdomen skin was 23.32 mGy and the minimum absorbed dose in the eye region was 0.15 mGy. The readout results are correlated with the results of spiral CT detector with the “ALARA” principle, we recommend suitable techniques should be selected to reduce absorbed dose of critical organ without reducing image quality.
Conclusion
More research is required to investigate whether modification of the parameters used during routine spiral CT scan can be limited absorbed dose of critical organ without a significant loss of image quality.
Language:
Persian
Published:
Journal of Advanced Biomedical Sciences, Volume:1 Issue: 3, 2012
Page:
75
magiran.com/p944020  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!