Background And Objective
Detection of microalbuminuria is an important laboratory test for evaluation of diabetic complications. A reliable method for determination of urine albumin is critical in this process. In the present study, we set-up and evaluated an electroimmunoassay (EIA) method for measearing the amount of albumin in urine. The reliability of Clinitek-100 Instrument was evaluated for screening of microalbuminuria as well.
Materials And Methods
In this study which was the evaluation of a diagnostic test, anti human albumin antiserum was prepared following the immunization of rabbit and the EIA method was set-up for albumin determination. The ratio of albumin to creatinine, as the index of albumin excretion rate, were calculated for 61 urine samples of diabetic patients. Finally, the reliability of EIA and clinitek-100 for detection of microalbuminuria were evaluated. An enzyme-linked immunosorbent assay (ELISA) was used as a reference method for determination of urine albumin and for detection of microalbuminuria as well.
Results
The detection limit for EIA method was 2 mg/L, with analytical range of 2-200 mg/L. The intra-assay coefficient variation (C.V) of various albumin concentrations were 3.34 to 5.57%, whereas the regression coefficient was 0.996 (r=0.996). The sensitivity and specificity of EIA-test for detection of microalbuminuria were 95% and 83%, respectively. The sensitivity and specificity of Clinitek-100 for microalbuminuria detection were 67% and 94%, respectively.
Conclusion
The EIA is a precise and specific method for determination of urine albumin and detection of microalbuminuria, but it is time consuming and unsuitable for single test reports. Clinitek-100 is specific and fast for delection of microalbuminuria. This method is suitable for single reports, but its negative results in test are controversial. We suggest that more care be taken for the clinitek-100 results and the negative results should be re-examined by other sensitive methods.
Language:
Persian
Published:
Journal of Rafsanjan University Of Medical Sciences, Volume:5 Issue: 4, 2007
Page:
225
magiran.com/p419095  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!